Chapter 17

Radiographic Quality

 

 

Radiographic quality is that feature of diagnostic radiology that describes how faithfully you have been able to reproduce shadows on the radiograph that clearly depict the anatomical organs under investigation.

17.1 Density, Contrast, and Detail

Radiographic quality is dependent on radiographic density, radiographic contrast, and geometric factors affecting radiographic detail(Table 17-1).

Radiographic density is determined by the number of photons that reach the film. This is influenced by beam quality as well as the type and thickness of tissue. Radiographic density is determined by the number of electrons present in a given thickness of tissue which influences the photon stopping power that the tissue possesses. The electrons may be placed so that the mass per unit volume is great or the number may be high because of the thickness of tissue. If the density or thickness of the tissue doubles, the number of photons reaching the film is halved and

Table 17-1

RADIOGRAPHIC QUALITY

-

Radiographic density

Radiographic contrast

Radiographic detail visibility

Production of x-rays

mA*

kVp*

exposure time*

wave form

beam filtration

incoming line voltage

kVp*

beam filtration

beam collimation

geometric unsharpness* focal spot size

Formation of X-ray image length

subject*

density

thickness

composition

use of contrast

media

focal-film distance

film speed

grid ratio

screen speed

beam attenuation

filters

grids

field size

subject contrast*

composition

thickness

use of contrast

media

fog or scatter

radiation*

secondary radiation

light fog

radiation fog

geometric unsharpness*

focal-film distance

object-film distance

film-screen contrast

 

patient thickness

motion unsharpness

(patient,tube,film)*

film speed

screen speed

subject thickness(anything that length-hens exposure time)

distortion

unequal enlargementof object

Processing

Processing features*

time

temperature

inherent film contrast

age of film

quality of film

processing features

time

temperature

chemistry age and

type

inherent unsharpness

film type(filmgraininess)

screen type (screenmottle

*(important factors)

    

RADIOGRAPHIC DENSITY.

Radiographic density is influenced by tissue density. Bone is almost twice as dense as muscle.

radiographic density is halved (Fig. 17-1). Therefore, radiographic density is inversely proportional to tissue density (Table 17-2).

Since radiographic density is a result of the number of photons that reach the film it is also influenced by:

Contrast is divided into radiographic contrast and subject contrast. Subject contrast is the most important factor in determination of radiographic contrast and it is helpful to introduce both terms at this time.

Radiographic contrast refers to the observed differences in radiographic density between adjacent areas in the radiograph. These differences in film exposure permit identification of shadows on the radiograph. Radiographic contrast depends on

Subject contrast is created by the difference in density and mass of two adjacent anatomic structures that permits transmission of x-ray photons through the two organs to a different degree. Subject contrast depends on the

The intensity of the transmitted beam is inversely proportional to the thickness of the part. If

the amount of tissue is doubled the intensity of the transmitted beam is cut to one half. The

Table 17-2

TISSUE DENSITY

Tissue

Density gm/cm3

Atomic number

Tissue density

Radiographic density

bone

1.85

13.8

High

Low

soft tissues,

-

-

-

-

water

1.00

7.42

-

-

fat

0.91

5.92

-

-

air

0.00129

7.64

Low

HIGH

    

   RADIOGRAPHIC CONTRAST.

Contrast is dependent on kVp range. Low kVp produces high radio graphic contrast while high kVp produces low radiographic contrast.

greater the density (mass per unit volume) the greater is the ability of the object to attenuate the beam. Photoelectric absorption is one of three methods of attenuation of the x-ray beam, and this form of absorption is increased in tissue with high atomic numbers. Bone will attenuate many more x-rays than muscle or fat assuming equal thickness.

The use of contrast media obviously affects subject contrast markedly. Positive contrast media contain iodine or barium, while negative contrast media are air, COz, or NOz The amount of contrast media plus their concentration determine the effect on tissue density. Use of a large quantity of undiluted barium sulfate in the stomach creates a much more dense shadow than use of a small amount of diluted iodinated contrast media in the shoulder joint of a dog. In all cases the purpose in using contrast media is to artificially alter subject density to influence subject contrast.

The type of x-ray film used affects the radiographic contrast produced by its inherent film contrast. Some types of film have a steep H and D curve (characteristic curve) and produce radiographs of high contrast. Other film has a more sloping curve to the H and D curve and the film produces radiographs of lower contrast.

The ability of an x-ray photon to penetrate tissue is also dependent on its energy. Low kVp settings produce high radiographic contrast provided the energy of the incident beam is great enough to penetrate the part to be examined. High kVp settings produce lower radiographic contrast referred to as long-scale contrast because there is a long scale of shades of gray with little difference between each separate shade (Fig. 17-2).

Any fogging or scatter radiation affects radiographic contrast because it decreases the density difference between two adjacent shadows, thus decreasing film contrast. Thus, film processing influences radiographic contrast greatly. The use of high temperature solutions, short developer times, and fast processing chemistries accentuates the affect on radiographic contrast.

Radiographic sharpness or detail is the feature of the radiograph that describes the definition of the edge of an anatomic structure. There are several types of unsharpness or loss of detail. Geometric unsharpness is due to

Motion unsharpness is due to motion of the tube, object or film during an exposure. Screen unsharpness or inherent unsharpness is due to the use of film and intensifying screens and is dependent on the speed and type of screen and film selected (Table 17-3). Absorption unsharpness is due to the shape of the structure radiographed. Round or oval objects have the greatest degree of unsharpness. Parallel unsharpness is due to the film emulsion being placed on both sides of the film base. This has a negligible effect on image unsharpness.

Another aspect of radiographic unsharpness is referred to as distortion. Distortion is the phenomenon in which there is enlargement of the various parts of the radiographic image by differing amounts. This will occur if the focal spot is not vertically above the object. The recording plane (film) may not be parallel to the plane of the object which causes different object-film distances and results in a distorted image.

If the object is separated from the plane of the film but the focal spot is directly overhead we appreciate equal distortion of the image and refer to this as magnification and may use it as an aid to diagnosis.

In the radiography of bones it is often possible to make the central ray nearly perpendicular to the anatomical structure and also have the bone parallel to the film. This results in near anatomical reproduction of the image on the radiograph. Compare that description to one of the position of the small bowel loops in the abdomen in which case the central ray may not be perpendicular to the bowel loop and the loop may be positioned at a large angle with the film. For this reason the radio graphic reproduction of bowel loops is often bizarre and causes great problems in interpretation.

Screen mottle. A method of measuring sharpness of image detail is to determine the line spread function. Light from an intensifying screen spreads causing blurring of an image edge. The line spread function is a method of measuring this effect.

A 10µ thick slit in metal placed in close contact with the film causes such marked collimation of the x-ray beam that it can be considered a line source x-ray beam.

RESOLUTION RESULTS OF SCREEN/FILM

COMBINATIONS.*

Screen/Film Combination

Resolution (line pairs/mm)

Calcium tungstate

-

Radelin T2/Cronex 6

5.7

Radelin UD/Cronex 6

6.7

Rare earth type

-

Trimax 2/XUD

9.6

Trimax 2/XDL

9.1

Trimax 2/XM

7.0

Trimax 4/XUD

9.5

Trimax 4/XDL

6.6

Trimax 4/XM

5.7

Trimax 8/XUD

8.0

Trimax 8/XDL

5.8

Trimax 8/XM

5.5

Trimax 12/XUD

7.5

Trimax 12/XDL

5.2

Trimax 12/XM

4.8

Lanex Fine/Ortho G

8.7

Lanex Fine/Ortho H

8.4

Lanex Regular/Ortho G

5.8

Lanex Regular/Ortho H

5.4

  *(Koblik, et al. 1980)

 

A perfect imaging system would produce an image on the developed film of a line 10µ thick. However, this does not happen because of light diffusion by the screens and the width of the exposed line on the film is wider extending several hundred microns to each side of the line. Measurement of the width of this line provides a method of expressing the faithfulness of the screens in production of a sharp image edge. The speed of the screens obviously affects the width of the exposed line with high speed screens causing a much wider line than slow speed screens.

Line spread function can also be used to measure the influence on the image of other factors such as:

Modulation transfer function is a concept formulated to provide an objective measurement of the combined effects of sharpness and resolution. The MTF represents a ratio between the information recorded and the total amount of information available. The MTF can never be greater than 1 since the information recorded can never exceed the available information. The information taken from determination of the line spread function is treated mathematically and presented on a graph when the percentage of information transferred to the film is determined for progressively larger number of line pairs per mm. The transfer of information at a low number of line pairs is better than that transferred at a high number of line pairs.

Radiographic mottle is defined to mean the impression gained by an observer when looking at the radiograph that there is marked graininess present which interferes with the interpretation of the radiograph. This is another cause of loss of radiographic detail. There are several components that create the course density fluctuations observed on screen films. The first is the density fluctuation that results from the x-ray photons that are absorbed in the intensifying screens. If the number of x-ray photons absorbed per unit area by the intensifying screen becomes low enough, adequate film density may be obtained but the statistical distribution of the x-ray photons prevents development of a sharp image on the film. This is called quantum mottle. Second, structural inhomogenities in the phosphor coating of the screens can cause a density fluctuation called structure mottle. The combination of these two phenomena is referred to as screen mottle. The third component is film graininess which is due to the built-in random distribution of exposed grains in the emulsion. Of these three, quantum mottle is the most significant (Rossman, 1963) (Table 17-4).

Factors that affect radiographic contrast usually are not so severe that they affect detail visibility. Graying of the radiograph from fog certainly affects the density of the parts of the film that would otherwise have been normally unexposed. Problems in geometry are the usual cause of loss of radiographic detail. These films do not have diminished contrast or density. However; if the production of scatter is so severe, it is possible that loss of detail may result as well as a diminished contrast. The following chart lists what are felt to be the factors most likely to cause change in density, contrast, or detail, recognizing that all three factors may be commonly affected by some technical errors.

17.2 Technical Errors in Film Handling and Film Processing

Technical errors due to exposure and film handling are found with both manual (wet tank) and automatic film processing. These various technical errors are encountered frequently in radiography and often destroy the diagnostic qualities of the radiograph. The failures arise from not appreciating correct radiographic procedures and radiographic processing and, therefore, incorrectly handling and processing a radiographic film. The following chart lists most common technical errors, their appearance on the radiograph, their cause, and method of correction .

Examination of the film with transmitted light and with reflected light often assists in determining the cause of the technical error. Any problem of too great or too little exposure of the film by visible light or radiation will result in an alteration in the number of silver crystals present on the film. This is detected by transmitted light. If the film is examined with reflected light, the film emulsion is noted to be intact and without damage or altered appearance. Errors that result from too rough handling of the film of ten damage the film emulsion. These may appear with transmitted light to be due to a lack or excess of silver crystals. However, reflected light may more clearly detect the damaged emulsion.

The method and frequency of film quality control obviously will be determined by the level of success you are having with radiographic techniques. The more problems, the more rigid must be the method of film quality control until you solve the difficulties. Remember that you have need for this control with both wet tank processing and with automatic processing (Tables 17-5,17-6).

One of the most common technical errors concerns the tendency for radiographs to turn yellow or yellow brown with age. The archival quality of the radiograph, or its ability to avoid this color change, is dependent on the residual ammonium thiosulfate that remains after the final wash. If the residual thiosulfate is greater than 25 ugm/in2, the film will turn yellow, and if the residual thiosulfate is greater than 100 ugm/in2, the film will turn brown within one year.

Table 17-5

TECHNICAL ERRORS (OTHER THAN PROCESSING).

Appearance on the Radiograph

Cause

Correction

1. Film density increased but contrast retains unless error is severe

Incorrect machine setting Too short focal-film distance Beam filter not in position Timer out of calibration over measurement of part to be examined

Surge in incoming line voltage Wrong type of screen/film combination

Lower kVp, mA, or time setting Correct focal-film distance Replace filter Repair timer

Remeasute part to examined

Evalute electrical circuit Check type of screen or film

2. Film density decreased

Incorrect machine settings Increased focal-film distance

One or more valve tubes burned out

Drop in incoming line voltage

Timer is out of calibration Under measurement of part to be examined

X-ray tube failure Increased beam-filtration Wrong type of screen film combination

Increase kVp, mA,or time setting

Deacrease focal-film distance replace valve tubes

Evaluate electrical circuit

Replace timer

Remeasure Part to be examined

Replace X-ray tube

Check filters

Check type of screen or film

3. Film gray with loss of contrast(fog)

Film exposed to radiation during examination

Film exposed to scatter radiation

Film exposed to radiation during storage

Films exposed to backscatter

radiation

Film stored under conditionsthat are too hot or too high humidity

Film old

Remove loaded cassettes from

room during examination

Use grid on tissue thickness

greater than 11 cm

Do not store film near source of

Use lead-backed cassettes and

radiation avoid placing cassette against a

hard surface

Store where dry and cool

Discard film

 4. Black marks or areas (not generalized)

Linear scratches on film Crescent marks due to bending film roughly or after exposure

Static electricity(linear dots or "tree" pattern) (Fig 17-3)

End of films black due to light leak while in film storage bin

Border of film black due to light leak because of poorly fitting cassette back

Handle film carefully

Handle film carefully

Handle film without causing friction; avoid low humidity in darkroom

Repair or replace film bin

Replace felt strip or replace damaged cassette

5. White artifacts (not generalized)

Fingerprints or blotches due to fixer one film prior to processing

"Hair" marks or irregular

marks due to dirt, debris,or chemistry stain on intensifying screen

Crescent marks due to bending film gently

usually prior to exposure

Irregular white marks due to dirt or medicament con-taining iodine on skin

White streaks near edge of film

Contrast medium on cassette, patient,or on table top

Clean bench surface in darkroom

and/or clean hands before handling film

Clean or replace intensifying screens

Handle film more carefully

Clean skin and hair coat prior to radiography

Cracks in screens

Clean cassette face or table top

6. Distorted, magnified, or blurred radiographic image

Blurred image due to patient motion

Blurred image due to motion

Blurred image due to (cassette) motion

Magnified image due toobject-film (cassette)

distance too great

Distorted image due to central beam not perpendicular to film (cassette)

Distorted image due to cen-tral beam not directed to center of film

Distorted image due to poor film-screen contact

Use compression device to

decrease tissue thickness and

permit shorter exposure time

tube Check tube locks or method of

stabilizing the tube

film Use sturdy cassette holders or

decrease exposure time

Position f closer to

anatomical part of interest

Reposition film(cassette) so central beam prependicular cassette to f(cassette)

Reposition tube so central beam hits center of film (cassette)

Clean foreign material from screen, correct damaged cassette or latch,

correct improperly mounted screens

7. Heavy lines across

Grid lines due to use of focal film distance outside range of grid focus

Grid lines heavy on one side

of film and more normal on

other side of film because grid not perpendicular to central beam

Grid liens heavy on one side

and more normal on other

because central beam not centered on midline of grid

Grid lines heavy on both edges of distorted due to grid damage

Grid lines randomly heavy

distorted due to grid damage

Use correct focal-film distance as radiograph determined by individual grid

Reposition so grid perpendicular to central beam

Reposition so central beam of film perpendicular to midline of grid

Turn grid over with primary beam hitting side indicated and Replace grid

8. Edge(s) of film white(underexposed)

Cone cut because of incorrect centering

Bucky tray not correctly positioned under table

Reposition so central beam hits center of cassette

Be certain to reposition bucky tray completely

9. Inconstant film density

Target damage

Replace x-ray tube

10. Double exposure

Cassette exposed twice without changing films

Position cassettes a parti- way to indicate exposure

Check fuses, check circuit breaker, check voltage selection for dead button, check interlock switch, check hand-foot switch selector, check tube over load protection indicator

11. Blank film

No exposure

Tube malalignment

Correctly center x-ray beam on film

Table 17-6

TECHNICAL ERRORS THAT RESULT FROM IMPROPER MANUAL (WET TANK) FILM PROCESSING

Most of these problems deal with rough handling of the film to be processed, failure to maintain correct solution temperature and proper strength of developer and fixer solution, and failure to move the film through the processing cycle at the prescribed time intervals.

Appearance on the Radiograph

Cause

Correction

1. Film density increased but contrast retains unless error is severe

Film overdeveloped due to too long a time in the developer

Film overdeveloped due high temperature of processing solution

Film overdeveloped due to inaccutrate thermometer

Film overdeveloped due to incorectly mixed developer

Corrct length of time in developer solution (5 min)

Correct temperature 20oC (68oF0)

Replace thermometer

2. Film density decreased

Film underdevelope due to too short time in developer

film undredeveloper due to low temperature of processing solutions

Film underdeveloped due to diluted, contaminated, or improperly mixed developer

Film underdeveloped deu to exhausted developer

Film underdeveloped due to inaccurate thermometer

Correct length of time in developer solution (5 min)

Correct temperature of processing solution (200C 9 68Fo)

Replace developer solution

Replace developer solution

Replace thermometer

3. Film gray with loss of contrast(fog)

Film exposed to rlight fog

 

Film affected by chemical fog dur to high temperature

Film affected by chemcal fog

Film overexposed and underdeveloped

correct defective darkroomsafety light or darkroom light leak

Do not view films intil they are completely cleared

Correct developing temperature, 20oC (68oF)

Discard old processing solutions

Corect exposure factors and develop film for 5 minute

 4. Black marks or areas (not generalized)

Black areas on two films Processed together due to film sticking togethier in fix and not clearing properly Developer solution on film

Gently agitate film in fix

Clean bench surface in darkroom, dry hands before loading film

5. Clear areas on film

Clear areas on two films processed together because films stuck together in the developer

Clean area on one film due to film being in contact with tank wall

Bubbles (air bell) due to nonagitation of film

Linear streaks where emulsion has been scratched away

Gently agitate films in the developer

Gently agitate films in the developer

Agitate film briefly after placing in developer

Careful exhausted fixer or fix film in processing tanks

6. Yellow radiation

Irregular yellowing of radiagraph with age due to improper fixation of the film

Replace exhausted fixer or fix film for correct time

7. Yellow patches

Film sticking togther (or to edge of tank)during

Agitate film gently during fixation; avoid crowding films in fix tank

8. Grit on film

Dirty water

Install filter in water line

Clean dryer

9. uneven development with lower half of film underdevelpled

Solution in bottom of tank is colder than on top

Developer was replenished

Stir solution prior to use

Stir solution prior to use

10. Streaks on film

Film hangers with developer on hanger clips

Water dropping on film

While drying

Washing film with dirty wash water

Stir solution prior to use

Clean film hangers

Use film dryer

Install filter on water line or increase flow rate in wash tank

11. Film cloudy and streaky (crusty or frosty when dry)

Dreaky with stichky feeling

Failure to fix for correct time interval

12. Brittleness of radio-graph

Excessive drying temperature; excessive drying time

Repair dryer

13. Reticulation Pattern

Uneven swelling and shrinking emulsion due to wide differ- ences in temperature of processing solutions

Control temperature of processing solutions

14. Corner marks

Wet or dirty fingers; dirty hanger clips

Dry hands before loading film and clean film hangers

15. Multiple dark spots

Developer on films prior to processing

Keep banch area clean

16. White artifacts

fixer on film prior to processing

Clean bench surface in darkroom, clean hands prior to loading film

 

Table 17-7

TECHNICAL ERRORS THAT RESULT FROM IMPROPER AUTOMATIC FILM PROCESSING

The automatic processor removes the chance of many of the errors created by the technician but creates some new ones. Errors in manual processing are often random in their distribution and do not follow specific patterns. Technical errors in automatic processors tend to recur until corrected since they are "machine generated. "Most of the problems relate to incorrect replenisher rates, incorrect temperatures, and problems in film transport.

Appearance on the Radiograph

Cause

Correction

1. Film density increased but contrast retains unless error is severe

Developer temperature high

Overreplenising

Light leak

Feddbell signals too quickly

Transport speed is low

Corrct temperature setting

Check replenisher pump and rate of replenisher

Reposition processor cover

Repair problem

2. Film density decreased

Developer temperature low Under replenshing

Transport speed too fast

Exhausted developer

Developer valve closed or line clogged

Microswitch malfunctioning

Air in developer line'

Replenishment tank too big,

Permitting solution oxidation

Corrct temperature setting Check replenisher pump and rete of replenishing

Repair Problem

Replace developer

Open valve or clean line

Check and adjust Bleed air

Correct tank size

3. Film scratches

Guide shoes malfuntioning if lines 2" apart the length of the film

Dryer air tubes not seated properly

Read just guide shoes

Repair

 4. Streaks on film

Top crossover rollers dry or crusted

Stangnant or diry wash water

Dirt air tubes

Feed bell out of phase and light turned on too soon

Clean rollers and run "Cleanup" film through processor

Correct problem

clean tubes

Correct feed bell

5. Black marks on film

Pressure marks: 1' apart -3"apart

Check small rollers

Check large crossover rollers

 

6. Damp or tacky films

Insufficient drying

Dryer temperature too low

Insufficient air venting

Insufficient fixation or hardening

Clogged air filter

Check thermostat

Check heater for correct operation

Check air venting system

Check fixer replenisher pump

7. Film overlapped

Film fed into processor too quickly

Roller tension is too tight inadequate wash-water flow

Adjust tension

Open and clean water mixing valves

Check supply pressure

Check strainer screens

8. Staining

Failure to wipe down cross over rollers

Clean rollers each night

    

FILM QUALITY COMPROMISED BY STATIC ELECTRICITY

 

References