|
|
|
|
|
|
Radiographic quality is that feature of diagnostic radiology that describes how faithfully you have been able to reproduce shadows on the radiograph that clearly depict the anatomical organs under investigation.
Radiographic quality is dependent on radiographic density, radiographic contrast, and geometric factors affecting radiographic detail(Table 17-1).
Radiographic density is determined by the number of photons that reach the film. This is influenced by beam quality as well as the type and thickness of tissue. Radiographic density is determined by the number of electrons present in a given thickness of tissue which influences the photon stopping power that the tissue possesses. The electrons may be placed so that the mass per unit volume is great or the number may be high because of the thickness of tissue. If the density or thickness of the tissue doubles, the number of photons reaching the film is halved and
|
|
|
Radiographic detail visibility |
Production of x-rays |
mA*kVp*exposure time*wave formbeam filtrationincoming line voltage |
kVp*beam filtrationbeam collimation |
geometric unsharpness* focal spot size |
Formation of X-ray image length |
subject*densitythicknesscompositionuse of contrastmediafocal-film distancefilm speedgrid ratioscreen speedbeam attenuationfiltersgridsfield size |
subject contrast*compositionthicknessuse of contrastmediafog or scatterradiation*secondary radiationlight fogradiation fog |
geometric unsharpness*focal-film distanceobject-film distancefilm-screen contrastpatient thicknessmotion unsharpness(patient,tube,film)*film speedscreen speedsubject thickness(anything that length-hens exposure time)distortionunequal enlargementof object |
Processing |
Processing features*timetemperature |
inherent film contrastage of filmquality of filmprocessing featurestimetemperaturechemistry age andtype |
inherent unsharpnessfilm type(filmgraininess)screen type (screenmottle |
![]() |
radiographic density is halved (Fig. 17-1). Therefore, radiographic density is inversely proportional to tissue density (Table 17-2).
Since radiographic density is a result of the number of photons that reach the film it is also influenced by:
Contrast is divided into radiographic contrast and subject contrast. Subject contrast is the most important factor in determination of radiographic contrast and it is helpful to introduce both terms at this time.
Radiographic contrast refers to the observed differences in radiographic density between adjacent areas in the radiograph. These differences in film exposure permit identification of shadows on the radiograph. Radiographic contrast depends on
Subject contrast is created by the difference in density and mass of two adjacent anatomic structures that permits transmission of x-ray photons through the two organs to a different degree. Subject contrast depends on the
The intensity of the transmitted beam is inversely proportional to the thickness of the part. If
the amount of tissue is doubled the intensity of the transmitted beam is cut to one half. The
Tissue |
Density gm/cm3 |
Atomic number |
Tissue density |
Radiographic density |
bone |
|
|
|
|
soft tissues, |
|
|
|
|
water |
|
|
|
|
fat |
|
|
|
|
air |
|
|
|
|

greater the density (mass per unit volume) the greater is the ability of the object to attenuate the beam. Photoelectric absorption is one of three methods of attenuation of the x-ray beam, and this form of absorption is increased in tissue with high atomic numbers. Bone will attenuate many more x-rays than muscle or fat assuming equal thickness.
The use of contrast media obviously affects subject contrast markedly. Positive contrast media contain iodine or barium, while negative contrast media are air, COz, or NOz The amount of contrast media plus their concentration determine the effect on tissue density. Use of a large quantity of undiluted barium sulfate in the stomach creates a much more dense shadow than use of a small amount of diluted iodinated contrast media in the shoulder joint of a dog. In all cases the purpose in using contrast media is to artificially alter subject density to influence subject contrast.
The type of x-ray film used affects the radiographic contrast produced by its inherent film contrast. Some types of film have a steep H and D curve (characteristic curve) and produce radiographs of high contrast. Other film has a more sloping curve to the H and D curve and the film produces radiographs of lower contrast.
The ability of an x-ray photon to penetrate tissue is also dependent on its energy. Low kVp settings produce high radiographic contrast provided the energy of the incident beam is great enough to penetrate the part to be examined. High kVp settings produce lower radiographic contrast referred to as long-scale contrast because there is a long scale of shades of gray with little difference between each separate shade (Fig. 17-2).
Any fogging or scatter radiation affects radiographic contrast because it decreases the density difference between two adjacent shadows, thus decreasing film contrast. Thus, film processing influences radiographic contrast greatly. The use of high temperature solutions, short developer times, and fast processing chemistries accentuates the affect on radiographic contrast.
Radiographic sharpness or detail is the feature of the radiograph that describes the definition of the edge of an anatomic structure. There are several types of unsharpness or loss of detail. Geometric unsharpness is due to
Motion unsharpness is due to motion of the tube, object or film during an exposure. Screen unsharpness or inherent unsharpness is due to the use of film and intensifying screens and is dependent on the speed and type of screen and film selected (Table 17-3). Absorption unsharpness is due to the shape of the structure radiographed. Round or oval objects have the greatest degree of unsharpness. Parallel unsharpness is due to the film emulsion being placed on both sides of the film base. This has a negligible effect on image unsharpness.
Another aspect of radiographic unsharpness is referred to as distortion. Distortion is the phenomenon in which there is enlargement of the various parts of the radiographic image by differing amounts. This will occur if the focal spot is not vertically above the object. The recording plane (film) may not be parallel to the plane of the object which causes different object-film distances and results in a distorted image.
If the object is separated from the plane of the film but the focal spot is directly overhead we appreciate equal distortion of the image and refer to this as magnification and may use it as an aid to diagnosis.
In the radiography of bones it is often possible to make the central ray nearly perpendicular to the anatomical structure and also have the bone parallel to the film. This results in near anatomical reproduction of the image on the radiograph. Compare that description to one of the position of the small bowel loops in the abdomen in which case the central ray may not be perpendicular to the bowel loop and the loop may be positioned at a large angle with the film. For this reason the radio graphic reproduction of bowel loops is often bizarre and causes great problems in interpretation.
Screen mottle. A method of measuring sharpness of image detail is to determine the line spread function. Light from an intensifying screen spreads causing blurring of an image edge. The line spread function is a method of measuring this effect.
A 10µ thick slit in metal placed in close contact with the film causes such marked collimation of the x-ray beam that it can be considered a line source x-ray beam.
|
|
Calcium tungstate |
|
Radelin T2/Cronex 6 |
|
Radelin UD/Cronex 6 |
|
Rare earth type |
|
Trimax 2/XUD |
|
Trimax 2/XDL |
|
Trimax 2/XM |
|
Trimax 4/XUD |
|
Trimax 4/XDL |
|
Trimax 4/XM |
|
Trimax 8/XUD |
|
Trimax 8/XDL |
|
Trimax 8/XM |
|
Trimax 12/XUD |
|
Trimax 12/XDL |
|
Trimax 12/XM |
|
Lanex Fine/Ortho G |
|
Lanex Fine/Ortho H |
|
Lanex Regular/Ortho G |
|
Lanex Regular/Ortho H |
|
A perfect imaging system would produce an image on the developed film of a line 10µ thick. However, this does not happen because of light diffusion by the screens and the width of the exposed line on the film is wider extending several hundred microns to each side of the line. Measurement of the width of this line provides a method of expressing the faithfulness of the screens in production of a sharp image edge. The speed of the screens obviously affects the width of the exposed line with high speed screens causing a much wider line than slow speed screens.
Line spread function can also be used to measure the influence on the image of other factors such as:
Modulation transfer function is a concept formulated to provide an objective measurement of the combined effects of sharpness and resolution. The MTF represents a ratio between the information recorded and the total amount of information available. The MTF can never be greater than 1 since the information recorded can never exceed the available information. The information taken from determination of the line spread function is treated mathematically and presented on a graph when the percentage of information transferred to the film is determined for progressively larger number of line pairs per mm. The transfer of information at a low number of line pairs is better than that transferred at a high number of line pairs.
Radiographic mottle is defined to mean the impression gained by an observer when looking at the radiograph that there is marked graininess present which interferes with the interpretation of the radiograph. This is another cause of loss of radiographic detail. There are several components that create the course density fluctuations observed on screen films. The first is the density fluctuation that results from the x-ray photons that are absorbed in the intensifying screens. If the number of x-ray photons absorbed per unit area by the intensifying screen becomes low enough, adequate film density may be obtained but the statistical distribution of the x-ray photons prevents development of a sharp image on the film. This is called quantum mottle. Second, structural inhomogenities in the phosphor coating of the screens can cause a density fluctuation called structure mottle. The combination of these two phenomena is referred to as screen mottle. The third component is film graininess which is due to the built-in random distribution of exposed grains in the emulsion. Of these three, quantum mottle is the most significant (Rossman, 1963) (Table 17-4).
Factors that affect radiographic contrast usually are not so severe that they affect detail visibility. Graying of the radiograph from fog certainly affects the density of the parts of the film that would otherwise have been normally unexposed. Problems in geometry are the usual cause of loss of radiographic detail. These films do not have diminished contrast or density. However; if the production of scatter is so severe, it is possible that loss of detail may result as well as a diminished contrast. The following chart lists what are felt to be the factors most likely to cause change in density, contrast, or detail, recognizing that all three factors may be commonly affected by some technical errors.
Technical errors due to exposure and film handling are found with both manual (wet tank) and automatic film processing. These various technical errors are encountered frequently in radiography and often destroy the diagnostic qualities of the radiograph. The failures arise from not appreciating correct radiographic procedures and radiographic processing and, therefore, incorrectly handling and processing a radiographic film. The following chart lists most common technical errors, their appearance on the radiograph, their cause, and method of correction .
Examination of the film with transmitted light and with reflected light often assists in determining the cause of the technical error. Any problem of too great or too little exposure of the film by visible light or radiation will result in an alteration in the number of silver crystals present on the film. This is detected by transmitted light. If the film is examined with reflected light, the film emulsion is noted to be intact and without damage or altered appearance. Errors that result from too rough handling of the film of ten damage the film emulsion. These may appear with transmitted light to be due to a lack or excess of silver crystals. However, reflected light may more clearly detect the damaged emulsion.
The method and frequency of film quality control obviously will be determined by the level of success you are having with radiographic techniques. The more problems, the more rigid must be the method of film quality control until you solve the difficulties. Remember that you have need for this control with both wet tank processing and with automatic processing (Tables 17-5,17-6).
One of the most common technical errors concerns the tendency for radiographs to turn yellow or yellow brown with age. The archival quality of the radiograph, or its ability to avoid this color change, is dependent on the residual ammonium thiosulfate that remains after the final wash. If the residual thiosulfate is greater than 25 ugm/in2, the film will turn yellow, and if the residual thiosulfate is greater than 100 ugm/in2, the film will turn brown within one year.
|
|
|
1. Film density increased but contrast retains unless error is severe |
Incorrect machine setting Too short focal-film distance Beam filter not in position Timer out of calibration over measurement of part to be examinedSurge in incoming line voltage Wrong type of screen/film combination |
Lower kVp, mA, or time setting Correct focal-film distance Replace filter Repair timerRemeasute part to examinedEvalute electrical circuit Check type of screen or film |
2. Film density decreased |
Incorrect machine settings Increased focal-film distanceOne or more valve tubes burned outDrop in incoming line voltageTimer is out of calibration Under measurement of part to be examinedX-ray tube failure Increased beam-filtration Wrong type of screen film combination |
Increase kVp, mA,or time settingDeacrease focal-film distance replace valve tubesEvaluate electrical circuitReplace timerRemeasure Part to be examinedReplace X-ray tubeCheck filtersCheck type of screen or film |
3. Film gray with loss of contrast(fog) |
Film exposed to radiation during examinationFilm exposed to scatter radiationFilm exposed to radiation during storageFilms exposed to backscatterradiationFilm stored under conditionsthat are too hot or too high humidityFilm old |
Remove loaded cassettes fromroom during examinationUse grid on tissue thicknessgreater than 11 cmDo not store film near source ofUse lead-backed cassettes andradiation avoid placing cassette against ahard surfaceStore where dry and coolDiscard film |
4. Black marks or areas (not generalized) |
Linear scratches on film Crescent marks due to bending film roughly or after exposureStatic electricity(linear dots or "tree" pattern) (Fig 17-3)End of films black due to light leak while in film storage binBorder of film black due to light leak because of poorly fitting cassette back |
Handle film carefullyHandle film carefullyHandle film without causing friction; avoid low humidity in darkroomRepair or replace film binReplace felt strip or replace damaged cassette |
5. White artifacts (not generalized) |
Fingerprints or blotches due to fixer one film prior to processing"Hair" marks or irregularmarks due to dirt, debris,or chemistry stain on intensifying screenCrescent marks due to bending film gentlyusually prior to exposureIrregular white marks due to dirt or medicament con-taining iodine on skinWhite streaks near edge of filmContrast medium on cassette, patient,or on table top |
Clean bench surface in darkroomand/or clean hands before handling filmClean or replace intensifying screensHandle film more carefullyClean skin and hair coat prior to radiographyCracks in screensClean cassette face or table top |
6. Distorted, magnified, or blurred radiographic image |
Blurred image due to patient motionBlurred image due to motionBlurred image due to (cassette) motionMagnified image due toobject-film (cassette)distance too greatDistorted image due to central beam not perpendicular to film (cassette)Distorted image due to cen-tral beam not directed to center of filmDistorted image due to poor film-screen contact |
Use compression device todecrease tissue thickness andpermit shorter exposure timetube Check tube locks or method ofstabilizing the tubefilm Use sturdy cassette holders ordecrease exposure timePosition f closer toanatomical part of interestReposition film(cassette) so central beam prependicular cassette to f(cassette)Reposition tube so central beam hits center of film (cassette)Clean foreign material from screen, correct damaged cassette or latch,correct improperly mounted screens |
7. Heavy lines across |
Grid lines due to use of focal film distance outside range of grid focusGrid lines heavy on one sideof film and more normal onother side of film because grid not perpendicular to central beamGrid liens heavy on one sideand more normal on otherbecause central beam not centered on midline of gridGrid lines heavy on both edges of distorted due to grid damageGrid lines randomly heavydistorted due to grid damage |
Use correct focal-film distance as radiograph determined by individual gridReposition so grid perpendicular to central beamReposition so central beam of film perpendicular to midline of gridTurn grid over with primary beam hitting side indicated and Replace grid |
8. Edge(s) of film white(underexposed) |
Cone cut because of incorrect centeringBucky tray not correctly positioned under table |
Reposition so central beam hits center of cassetteBe certain to reposition bucky tray completely |
9. Inconstant film density |
Target damage |
Replace x-ray tube |
10. Double exposure |
Cassette exposed twice without changing films |
Position cassettes a parti- way to indicate exposureCheck fuses, check circuit breaker, check voltage selection for dead button, check interlock switch, check hand-foot switch selector, check tube over load protection indicator |
11. Blank film |
No exposureTube malalignment |
Correctly center x-ray beam on film |
|
|
|
1. Film density increased but contrast retains unless error is severe |
Film overdeveloped due to too long a time in the developerFilm overdeveloped due high temperature of processing solutionFilm overdeveloped due to inaccutrate thermometerFilm overdeveloped due to incorectly mixed developer |
Corrct length of time in developer solution (5 min)Correct temperature 20oC (68oF0)Replace thermometer |
2. Film density decreased |
Film underdevelope due to too short time in developerfilm undredeveloper due to low temperature of processing solutionsFilm underdeveloped due to diluted, contaminated, or improperly mixed developerFilm underdeveloped deu to exhausted developerFilm underdeveloped due to inaccurate thermometer |
Correct length of time in developer solution (5 min)Correct temperature of processing solution (200C 9 68Fo)Replace developer solutionReplace developer solutionReplace thermometer |
3. Film gray with loss of contrast(fog) |
Film exposed to rlight fog
Film affected by chemical fog dur to high temperatureFilm affected by chemcal fogFilm overexposed and underdeveloped |
correct defective darkroomsafety light or darkroom light leakDo not view films intil they are completely clearedCorrect developing temperature, 20oC (68oF)Discard old processing solutionsCorect exposure factors and develop film for 5 minute |
4. Black marks or areas (not generalized) |
Black areas on two films Processed together due to film sticking togethier in fix and not clearing properly Developer solution on film |
Gently agitate film in fixClean bench surface in darkroom, dry hands before loading film |
5. Clear areas on film |
Clear areas on two films processed together because films stuck together in the developerClean area on one film due to film being in contact with tank wallBubbles (air bell) due to nonagitation of filmLinear streaks where emulsion has been scratched away |
Gently agitate films in the developerGently agitate films in the developerAgitate film briefly after placing in developerCareful exhausted fixer or fix film in processing tanks |
6. Yellow radiation |
Irregular yellowing of radiagraph with age due to improper fixation of the film |
Replace exhausted fixer or fix film for correct time |
7. Yellow patches |
Film sticking togther (or to edge of tank)during |
Agitate film gently during fixation; avoid crowding films in fix tank |
8. Grit on film |
Dirty water |
Install filter in water lineClean dryer |
9. uneven development with lower half of film underdevelpled |
Solution in bottom of tank is colder than on topDeveloper was replenished |
Stir solution prior to useStir solution prior to use |
10. Streaks on film |
Film hangers with developer on hanger clipsWater dropping on filmWhile dryingWashing film with dirty wash water |
Stir solution prior to useClean film hangersUse film dryerInstall filter on water line or increase flow rate in wash tank |
11. Film cloudy and streaky (crusty or frosty when dry) |
Dreaky with stichky feeling |
Failure to fix for correct time interval |
12. Brittleness of radio-graph |
Excessive drying temperature; excessive drying time |
Repair dryer |
13. Reticulation Pattern |
Uneven swelling and shrinking emulsion due to wide differ- ences in temperature of processing solutions |
Control temperature of processing solutions |
14. Corner marks |
Wet or dirty fingers; dirty hanger clips |
Dry hands before loading film and clean film hangers |
15. Multiple dark spots |
Developer on films prior to processing |
Keep banch area clean |
16. White artifacts |
fixer on film prior to processing |
Clean bench surface in darkroom, clean hands prior to loading film |
|
|
|
1. Film density increased but contrast retains unless error is severe |
Developer temperature highOverreplenisingLight leakFeddbell signals too quicklyTransport speed is low |
Corrct temperature settingCheck replenisher pump and rate of replenisherReposition processor coverRepair problem |
2. Film density decreased |
Developer temperature low Under replenshingTransport speed too fastExhausted developerDeveloper valve closed or line cloggedMicroswitch malfunctioningAir in developer line'Replenishment tank too big,Permitting solution oxidation |
Corrct temperature setting Check replenisher pump and rete of replenishingRepair ProblemReplace developerOpen valve or clean lineCheck and adjust Bleed airCorrect tank size |
3. Film scratches |
Guide shoes malfuntioning if lines 2" apart the length of the filmDryer air tubes not seated properly |
Read just guide shoesRepair |
4. Streaks on film |
Top crossover rollers dry or crustedStangnant or diry wash waterDirt air tubesFeed bell out of phase and light turned on too soon |
Clean rollers and run "Cleanup" film through processorCorrect problemclean tubesCorrect feed bell |
5. Black marks on film |
Pressure marks: 1' apart -3"apart |
Check small rollersCheck large crossover rollers
|
6. Damp or tacky films |
Insufficient dryingDryer temperature too lowInsufficient air ventingInsufficient fixation or hardeningClogged air filter |
Check thermostatCheck heater for correct operationCheck air venting systemCheck fixer replenisher pump |
7. Film overlapped |
Film fed into processor too quicklyRoller tension is too tight inadequate wash-water flow |
Adjust tensionOpen and clean water mixing valvesCheck supply pressureCheck strainer screens |
8. Staining |
Failure to wipe down cross over rollers |
Clean rollers each night |