Chapter 5

Fracture Treatment & Other Related Orthopedic Principles

 

 

 5.1 Fracture:

A complete/incomplete break in the continuity of bone or cartilage or both.

 5.2 Mechanics of Fractures

When sufficient force is applied to an object, it deforms the object. Deformation is measured by a force deformation curve.

Bone is not a static module

Simply stated: Bone is deposited where needed in response to mechanical stresses 

5.3 Forces on a bone:

Bone undergoes specific failure or fracture associated with different modes of loading when loaded in: 

5.3.1 Tension:

  • Fr plane is approx perpendicular to the applied force  Fr plane approx
  • 1 to applied force 

5.3.2 Compression:

  • Fr plane is generally an oblique angle to applied force 

5.3.3 Torsional:

  • Fr is in a spiral plane 

5.3.4 Bending Loads

(High tensile stress on one side & high compressive stress on other): Fr plane originates transversely on tension side with oblique. Fr plane on compressive side. If more than one oblique, Fr plane exists on compressive side, a "butterfly" segment results. 

5.3.5 Bending & Compression Loads

Acentuation of oblique Fr of bending loads with bigger butterfly segments or comminution. The later is due to a complex rapidly applied force. 

5.4 Causes of Fractures:

5.5 CLASSIFICATION OF FRACTURES

5.5.1 DIRECTION & LOCATION OF FRACTURE:

  • a) Transverse Fr: Fr line is at right angle to bone axis
  • b) Oblique Fr: Line of Fr diag. to long axis of bone. Fragments tend to slip.
  • c) Spiral Fr: Fr. Line is a curve
  • d) Comminuted Fr: splintering or fragmentation of bone. Fr lines meet at a common point.
  • e) Multiple/Segmental Fr: Bone broken into 3-4 more segments which don’t meet at a common point.
  • f) Impacted Fr: Segments driven firmly together
  • g) Avulsion Fr: A fragment of bone, which is the site of insertion of a muscle, tendon or ligament, is detached as a result of a forceful pull
  • h) Physeal Fr: Fr/Separation occurs at physeal line or growth plate. Seen in young animals.
  • i) Condylar Fr: Fr line passes through A condyle.
  • j) Intercondylar Fr: Fr line runs between two condyles

5.5.2 PRESENCE OF A COMMUNICATION EXTERNAL WOUND:

  • a) Closed fr: Does not communicate to outside
  • b) Open fr: Communicate to outside with contamination/infection 

5.5.3 EXTENT OF DAMAGE:

  • a)Complete Fr: Total disruption of bone usually with displacement.
  • b) Green stick Fr: One side of bone is broken and other is bent. Seen in young animals. Displacement is minimal.
  • c) Fissure Fr: One or More fine cracks penetrate the cortex in spiral or longitudinal direction. Periosteum is still intact

5.5.4 STABILITY FOLLOWING ANATOMICAL REDUCTION:

  • a) Stable Fr: FraGments interlock and resist shortening forces. eg. transverse, GreensticK, impacted.
  • b) UnstaBle Fr: Fragments slide by each other out of position. eg. oblique, spiral 

5.5.5 CLASSIFICATION OF OPEN FR:

  • a) 1st DeGree: occurinG from within, with Fr bone penetrating the sKin at the time of injury
  • b) Second Degree: occuring from outside, with contusion of the skin and soft tissues
  • c) Third Degree: occuring from outside with extensive skin, nuscle and possible nerve damage with a comminuted fracture. Damage is potentially difficult to treat. 

5.5.6 PHYSEAL INJURIES, CLASSIFICATION OF:

  • a) Salter I: occurs transversely through the region of cartilage hypertrophy and degeneration.
  • b) Salter II: extends to metaphysis
  • c) Salter III: extends from hypertrophied cartilage through germinal cell layer into epiphysis
  • d) Salter IV: Longitudinal Fr through metaphysis,physis & epiphysis into the joint
  • e) Salter V: Fr crushes the chondroblastic cell layer 

    5.5.6.1 Diagnosis of fractures:

    • - GPE
    • - Orthopedic Exam. i.e. History, clinical signs, radiograms etc.
    • Clinical signs
    • - Pain or tenderness
    • - Deformity
    • - Abn. mobility
    • - Local swelling
    • - Loss of function
    • - Crepitence 

5.6 BONE HEALING

5.7 PRIMARY FRACTURE HEALING:

Areas of osteogenic potential

Primary Fr. Healing occurs provided Fr ends are:

 5.8 FACTORS INFLUENCING BONE HEALING:

5.8.1 Local Factors:

5.8.1.1 Degree of local soft tissue trauma

  • - more trauma, retarded bone healing
  • - as mesenchymal cells are utilized to repair soft tissue damage
  • - Hematoma diffuses into adjacent soft tissues
  • - Extensive disruption of blood supply so increased devitalized tissue

5.8.1.2 Degree of bone loss:

  • - Loss of bone substance or excessive distraction Oc fragments leads to compromised ability of the cells to bridge the gap

5.8.1.3 Type of bone involved:

  • - In cancellous bone repair is rapid due to many points of bone contact which are rich in blood supply & cells
  • - Cortical bone heals through primary or standard secondary healing through ext. callus formation.

5.8.1.4 Degree of Immobilization:

  • - Inadequate mobility control leads to delayed or non union as motion disrupts the initial fibrin scaffoldino and external callus fails to form
  • - O2 tension & acidosis

5.8.1.5 Infection:

  • - If infection is superimposed on a fracture the local forces are mobilized to wall off & eliminate the infection, hence healing is retarded or is absent

5.8.1.6 Avascular necrosis:

  • - If one fracture segment is avascular healing is retarded or absent
  • -- If both fragments are avascular healing is very poor or mostly absent 

5.8.2 Systemic Factors:

5.8.2.1 Age:

  • - elderly mature animals heal at a slower rate than younger animals

5.8.2.2 Hormones:

  • - corticosteroids inhibit the union where as growth hormones stimulate fracture healing
  • - Thyroid hormones, calcitonin Insulin, vitamin A&D; anabolic steroids are reported to enhance bone healing
  • - Where as Diabetes, castration hypervitaminosis A & D and ractitic state retard bone healing

5.8.2.3 Exercise & Local stress:

  • - Denervation retards bone healing probably by diminishing stress across fracture line
  • - Exercise increases the rate of repair of bone 

5.9 HEALING RELATIVE TO DIFFERENT KINDS OF STABILITY FOLLOWING MIDSHAFT EXPERIMENTAL OSTEOTOMIES OF FEMUR 

5.9.1 No fixation:

  • - There was marked overriding
  • - Union When occurred was almost entirely dependent upon the organization of the fracture hematoma.
  • - Hematoma was huge & surrounded the fracture ends
  • - Hematoma differentiated into cartilage which attempted to form a bridging callus which when completed was slowly converted to bone by endochondral ossification
  • - Delayed union & non-unions were very common
  • - The earliest union occurred at 14 weeks posttruama or Fr. 

5.9.2 Loose fitting Intramedullary fixation:

  • - Rotation of fragments around the fixator
  • - Marked periosteal reaction extending the full length of the diaphysis on both fragments to within one cm. of the fracture line where periosteum has been traumatized
  • - Cortex beneath the periosteal reaction under went extensive remodeling and eventually by 3 weeks resembled a cancellous structure
  • - Presence of I.M nail inhibited the formation of endosteal callus
  • - Bone formation at Fr. site took place by organization of hematoma and endochondral ossification
  • - When motion was more pronounced the cartilaginous bridge never completed. A cleft developed at the level of fracture extended in soft tissue which in older specimen was lined by synovial like tissue complete with villi
  • - Nonunion or delayed unions ensued

5.9.3 Tight fitting I.M. nails:

  • - All rotary motion was eliminated
  • - Periosteal reaction was limited to 2-3 cm of the fracture ends
  • - No endosteal callus formed
  • - Bone formation occurred in fibrous stroma of the hematoma without going through cartilage phase
  • - A very little endochondral ossification takes place
  • - This type of healing may be seen in cases of stack pinning simple I.M. pins along with Krischner apparatus 

5.9.4 Bone Plate: no compression versus compression:

  • - Minimal periosteal reaction in compressed
  • - More periosteal reaction in non-compressed
  • - Marked endosteal reaction at three weeks and accounted for early union
  • - Primary healing from cortical elements took place

Relatively very little cartilage was found in medullary cavity callus of the compressed fracture

Mechanial tests show that fracture uniting by closed reduction and large periosteal callus by the end of first month would provide superior stiffness & strength than to those fractures compressed by tension band plates

  • - However tension band plates approach the same strength & stiffness of the above group in 3 months.
  • - The mode of osteogenesis is dependent on the stability of reduction of the bone fragments. Periosteal callus is directly proportional to instability. 

5.10 DELAYED UNION & NON-UNION

5.10.1 Delayed Union:

  • -When a fracture does not heal in an arbitrary normal healing time
  • - The normal time may vary according to age, species, breed, bone involved, level of fracture & type of fracture.
  • - Radiographically the Fr line is visible, there is minimal callus formation & Fr. site has a feathery or woolly appearance. 

5.10.2 Non-Unions:

  • - An un-united Fr. where repair process has totally ceased
  • - No osteogenic activity present
  • - It is one end or delayed union
  • - Radiographically Fr. Line is visible with no callus formation or union 

5.11 LASSIFICATION: 

5.11.1 VIABLE: i.e. capable of biological reaction

i) Hypertrophic or elephants foot from

  • - There is abundance of callus with a profusion of B.V.
  • -- Fr. Gap is occupied by fibro cartilage
  • - Cause is usually insufficient stabilization or premature weight bearing

ii) Slightly Hypertrophic or Horse hoof form:

  • - Milder form of hypertrophic non-union with poor callus or slight increased density of the fragment ends
  • - Frequently occurs under a plate fixator where Fr is some what unstable. Callus & plate are not sufficient to stop motion

Metal or screws fail and become loose

Oligotrophic Non Union:

  • - No callus formed
  • - Results from great displacement of fragments too much traction or lack of anatomical reduction or apposition
  • - Fr. ends round off, become shorter through resorption c& decalcify
  • - May be joined by fibrous tissue 

5.11.2 NON-VIABLE: i.e. incapable of biologic reaction

i) Dystrophic non union:

  • - Presence of an intermediate bony fragment, with conpromized blood supply, which has healed to one side but not with other main bony fragment.

ii) Necrotic Non-Union:

  • - Presence of numerous intermediate fragments with insufficient blood supply
  • - Fragments are more radiopaque
  • - Maybe further away from main segments

iii) Defect:

  • A segment of bone is lost or missing

iv) Atrophic

  • - End result of top three
  • - Fragment ends resorb inactivity leads to osteoporosis atrophy 

    Incidence of Non Union:

    • - About 5% in human surgery
    • - Less in animals
    • - Radius & Ulna 60%
    • - Tibia 25%
    • - Femur 15% 

5.12 CAUSES OF DELAYED/NON UNION

a. advised open surgery

5.13 PATHOGENESIS:

5.14 CLINICAL SIGNS:

5.15 DIAGNOSIS OF DELAYED/ NON UNION:

Clinical signs, History & Radiographs 

5.16 TREATMENT:

Aim is to improve the local physiological and mechanical environment so as to allow fracture healing to proceed.

5.16.1 TRADITIONAL THERAPY:

5.16.1.1 Compression without bone graft:

  • - When adequate reduction and alignment of the fracture ends were achieved initially and some evidence of callus at fracture site is present, compression (platino) will achieve healing within 6-12 wks. 

5.16.1.2 Compression with bone graft:

  • - When during initial repair alignment & fixation was inadequate then fracture site is scraped off, of the fibrous & cartilage components. Bones are anatomically aligned & seeded with cancellous bone graft & immobilized. Healing will ensue.

5.16.1.3 Electrical Stimulation Therapy:

  • - Direct electrical stimulation with an electrode placed into the non union site as well as non-invasive technique using electromagnetic fields & capacitive coupling are new modes of therapy .
  • - Bone resorption occursat-anode (+electrode)
  • - Bone deposition occurs at cathode (-electrode)
  • - Current levels used are 5 RA to 20 UA <5 UA fail to deposit bone >20 UA cause bone necrosis
  • - Three major devices approved by FDA for use in human osteogenesis are:
    • i) Direct current stimulator (Zimmer In)
    • ii) Inductive coupling (Electro-Biology N.J.)
    • iii) Direct current; completely implantable stimulator (Electronics Prop. LTD. Wis)
  • - Overall success rate of 80-37% has been reported for treatment of non-unions is man 

5.17 BONE GRAFTING/TRANSPLANTATION

5.19 ORTHOPEDIC IMPLANTS & BIOMATERIALS

Chemical Composition of 316 L

Carbon

0.030 % Max

Copper - 005%/i-1ax

Mn

2 0% Max

Iron - balance

Phosphorus

0.025% MAX

-

Sulfur

0.010 Max

-

Silian

0.75% Max

-

Cr

17-19%

-

NI

12 - 14%

-

Molybdenum

2 - 3%

-

N

O-10 Max

-

Copper

0.5% Max

-

Iron

balance

-

The finished product or implant is given a shine and protected by a layer of oxide (nitric) that acts as an electric resistor to retard the anodic dissolution by metal cations and hence reduces corrosion

Corrosion occurs due to metal transfer from orthopedic tools & equipment to the implant, destruction of oxide layer, mixing of various metals in the alloy leading to galvanic reaction, corrosion and other surface irregularities like pits or crevices.

Thus metallic devices for fracture fixation should be removed whenever the implant has done its job e.g. clinical healing has taken place.

Implant removal:

AGE

EXT. SKELT I.M. PIN

BONE PLATE

<3M

2-3 wks

4wks

3-6M

4-6 wks

2-3M

6-12M

5-8 wks

3-5M

>lyr

7-12 wks

5M-1yr

INTERACTION BETWEEN IMPLANT & BODY IN FR. RX

 Implant conditions

Probable effecton tissue

and Implant

Implant too Weak

Non union Pseudoarthiosis

Fatigue failure

Implant too rigid

Rarefaction of Cortex

Loosening OF lmpl .

Unstable fixation

Bone resorption Delayed healing

Fatigue

FailureFatigue Failure

Local relative motion

Tissue impregnation

Change in pll & O2 depletion

Fretting

-

-

Fretting

-

corrosion enhancedcollision 

5.20 PRINCIPLES OF FRACTURE RX

Pre-requists:

Objectives:

5.21 ORTHOPEDIC EQUIPMENT

Besides general instrument pack (basic) the orthopedic equipment often required is:

5.21.1 General:

  • - Retractors: Army - Navy, Rake, Senn. Hohman, Gelpi, Weitlaner
  • - Bone holding Forceps: Richards bone clamp, Ker Verbruggee & reduction
  • - Hand drills: Steinmann pin chuck, Rotary drill
  • - Bone rongeurs & cutter, currettes
  • - Osteotome & chisel & mallet
  • - Periosteal elevators
  • - Wire pliers, forceps & twister, cutters
  • - Pin cutters

5.21.2 Pinning equipment

  • - Pins assorted sizes: 3/8", 1/4",3/16",5/32",9/64", 1/8,7/64", 3/32", 5/64", .062", .045", .035".
  • - Pin vise/chuck
  • - Pin cutter

5.21.3 Wiring equipment:

  • - Orthopedic wire: 18_ 20, 22, gauze
  • -- Wire twister /tightener
  • - Wire pliers & cutter
  • - Wire passer
  • - Drill & pins to drill holes

5.21.4 Plating & screw equipment:

  • - Assortment of plates: eg. ASIF 4.5_ 3.5 2.7 systems
  • - Assorted screws: eg. ASIF 4.5;3.5, 2.7; cortical 4.0 & 6.5 mm cancellous
  • - Assorted washer & nuts
  • - Power drill or hand drill
  • - Drill bits: eg. 4 5 screw -->3.2 bit; 3.5 screw ---> 2.0 2 7 screw, -> 2.0 bit
  • - Drill guide: eg. DCP: > Neutral (green) Compression (yellow)
  • - Depth gauzes
  • - Bone Taps for non-self tapping screws
  • - Tan sleeves
  • - Screw drivers
  • - Templates
  • - Countersinks
  • - Plate bender &Torsion bar

5.21.4 Orthopedic Companies:

Richards Mgf. Memphis TN; Kirschner, Aberdeen Maryland; Synthesis Wayne, PA; Zimmer, Warsaw, IN; OEC, Bourbon,IN

5.22 FRACTURE TREATMENT

5.22.1 BONE SPLINTING PRINCIPLES:

A. Ext. Pin splints:

  • - Dr. Stader designed the first veterinary ext. pin splint in mid 30's.
  • - Most common ext pin splint was designed by Dr. Ehmet in the late 40's and is marketed by Kirschner Mfg. Co. as the K-E splint.

A1. K-E Splint components:

  • - Fixation pins: Trocar pointed steinmann pins, usually non-threaded. Single point small pins: 5/66" (1.95 mm) Diameter: Medium pins: 3/32" (2.38 mm); 1/8" (3-17 mm), Large Pins: 3/16 " (4.78 mm); 1/4 (6.35 mm)

A2. Ext. Fixation Rod/bar or connecting rod/bar

  • Small Rod: 1/8" (3.17 mm) diameter
  • Medium: 3/16" (4.72 mm) diameter
  • Large Rod: 7/16" (11.02 mm) diameter

A3. Fixation Clamps

Single: Small, Medium, Large

Double: Small, Medium, Large

  • - Krischmer Co. Manufactures and sells them as small, medium & large sets
  • - These clamps do accept regular Steinmann pins as fixation pins & connecting bars.

5.23 CLASSIFICATION OF K-E APPARATUS (Heirholzer)

5.23.1 Depending upon configuration:

  • Type I: Half pin: fixation pins pass through one skin surface & both cortices. Connecting bar(s) on one side
  • Type II: Full pin/thru-thru: fixation pins pass through both skin surfaces & both cortices, connecting bars placed on either (opposing) sides
  • Type III: Combination of type I, II: i.e Half & full pin each. Each system is interconnected to creat-three-Dimensional or a triangular frame

Detailed configuration can be devised with one's imagination. Some time a configuration may be devised and described along with the 'type', eg.

Type I: 2 to 3 pins: single bar

  • 4 to 6 pins: single bar
  • # of pins: double bar
  • # of pins: Double clamp, original tech.

Type II: 2-6 pins, standard technique

  • modified, 2 full pins, 2-4 half pins

Type III: 3 dimensional standard

  • # of pins thru-thru; # in half pin

5.23.2 As indications:

  • l) Adjunct to other internal fixation: to prevent axial rotation or/and collapse of fracture site eg. IM pin, wire, cross pin etc.
    • - can be removed in 3-5 wks ie. when callus becomes strong or organized to prevent rotation or collapse.
  • 2) Primary fixation: in Hypertrophic non-unions (in closed blind manner)
    • - Atrophic non -unions open reduction manner with decortification & bone grafting
    • - In severe communitions when exacting fixation is not possible in closed reduction manner
    • - open gunshot or infected fractures
    • - corrective osteotomies
    • - stable/non-stable fractures
    • - mandibular fractures
    • - Type II & III only below the stifle &elbow.
  • 3) Arthrodesis along with cross pinning
  • 4) Leg lengthening procedures

5.23.3 Advantages:

  • i) Atraumatic: does not invade or occupy the fracture (war zone) area
  • ii) Can be used closed
  • iii) Simple & quick
  • iv) Versatile alone or with other fixators eg. configuration
    • - 2 lag screws & pin (IM) cannot work whereas
    • - 2 screws & KE can
    • - IM Pin & KE can
  • v) Adequate skeletal fix without the physical presence of metal at Fr. site

5.23.4 Disadvantages:

  • l. Strength of fixation - not as great as with certain other devices.
    • - So use with caution on femorals & Humoral Fr in large dogs
  • 2. External Device: Infection around pins
    • - Requires additional care of the protruding appliance

5.23.5 Principles of Application:

  • - Asepsis
  • - Normal anatomical reduction. Maynot be perfect during closed application/reduction
  • - Pins must be maximally spaced in each segment
  • - Reduce pin group separation
  • - Minimize pin length
  • - Minimize length of connecting bars
  • - Pins must penetrate both bone cortices
  • - Clamps as close to skin as possible
  • - Use adequate size pins (not more than 30% the diameter of the bone) & connecting bars
  • - Insert without causing thermal necrosis
  • - No distortion of skin or underlying muscles during insertion
  • - 3 pins in each segment given 50 to 100% more strength in compression, bending & torsion than 2 pins.
  • - More than 4 pins per segment gives insignificant advantage in above (stiffness) terms.
  • - Most common technique of application is type I & or type II.

5.23.6 Application of type I Half-pin-splints:

Standard Procedure (4 pins, 2 short bars & conn bar):

  • - Paired pins are inserted in each segment at an angle of 35-450 .
  • - Each pair is connected with a short-bar one double clamp & 2 single clamps
  • - Connect the 2 single bars through the double clamps to a connecting bar

Advantage: Requires least amount of sophistication in pin placement but is bulky

  • - To increase strength 2 connecting bars maybe used

Brinker's Modification:

  • - Reduce the fracture (open or closed)
  • - Miaintain in reduce position during insertion of fixation pins
  • - Inset minimum of 2 pins in each fragment at same plane
  • - Pins must cross one skin & 2 cortices and must be inserted at 35-45 to each other
  • - Place the proximal (l) & Distal (4) pins relatively near the end of the bone.
  • - Attach the connecting bar with four clamps
  • - Tighten the clamps on the two end pins
  • - Insert the center 2 pins through clamps & tighten all clamps
  • - The splint is covered with gauze & tape
  • - In open reduced fracture splint is applied away from incision line.

Advantage Over Standard Procedure:

  • -- Aligns all 4 pins in one plane & connected by one single connecting bar
  • - Hence mechanically more stable with less tendency to loosen or twist (as connecting bar anchored at 4 places) than standard method where connecting bar attached at only one point in each segment (2 places total)
  • - Less equipment; bulk & cost etc.
  • - Provides elastic rather than rigid skeletal fixation. Hence this controlled mobility provides enhanced clinical fracture healing through exuberant periosteal callus formation.

5.23.7 Type II (Full pin splintage) Standard Type II:

  • - All fixation pins are driven through both skin surfaces (skin bone-skin) so that fixation bars can be placed on both (opposing) sides of the bone.
  • - Provides better support than type I i.e. 20-50% more stiffness over type I
  • - Fixation pins are driven perpendicular to the bone or at slight angle
  • - Later prevents medial to lateral motion of the fixator. It can also be accomplished by compressing the top 2 pins & bottom 2 pins together (toward each other) while tightening the fixation clamps
  • - Application limited to the bones below the stifle & elbow, primarily tibia & radius

Modified Type II:

  • - As it can be difficult to get all 4 fixation pins in the same plane on either side of the bone, many surgeon's modify the type II by only using one full pin in each segment with additional half pins for adequate stability.
  • - Same plane of pin insertion on either side of the bone can be achieved by using a pin guide or 2 connecting bars on one side

5.23.8 COMPLICATION OF K-E APPARATUS

  • i) Pin tract sequestra: Generally due to thermal necrosis if present should be removed
  • ii) Local Infections of pin tract & drainage: most common problem. Could be caused by
    • - excessive skin & soft tissue movement
    • - tension against the skin
    • - Loosening of the pins

Generally clears up if pin is tightly seated otherwise remove the loose pin

  • iii) Loosening of skeletal fixator: pins loosen
    • - due to angle of insertion being less than 35
    • - due to loosening of single fixation clamp
  • iv) Bending of the transfixation pins: due to
    • - too much weight bearing
    • - Insufficient pins (few in #)
    • - Inappropriate size
  • v) Iatrogenic fractures due to
    • - use of oversized pins
    • - pins placed too close together or near the fracture line (war zone) or in a fissure
    • - Unrestricted activity causing fracture through pin holes

5.23.9 CLIENT EDUCATION:

  • - Restrict activity
  • - Leash walk
  • - Avoid fencing etc which may catch or tangle the apparatus
  • - Inspect apparatus daily for any foul smell discharge etc
  • - Return every 2 wk s for checkups & evaluation
  • - Restricted activity to continue for additional 6-8 wks even following removal of splintage

5.23.10 PIN REMOVAL:

  • - Sedation/Anesthesia
  • - Disassemble the frame & extra pins by hand or chuck or needle
  • - nose vise grip pliers
  • - Remember to unscrew if threaded pins were used
  • - A small amount of serosanguineous fluid often drains from the pin site.
  • - A soft padded bandage may be used for a day or two

5.23.11 INTRAMEDULLARY PINNING

  • - It is the most readily used system of Internal fixation
  • - Devices used in veterinary medicine are Steinmann pins, K - wire Rush pin & Kuntschner nail. Last 2 are not used very frequently.

5.23.12 STEINMANN PINS & I.M. FIXATION

  • - Steinmann pins commonly used are 9" to 12" long and 1/16" to 1/1" in diameter.
  • - The pins maybe available in chisel Trocar & Threaded tlocar point.
  • - Maybe double pointed or single pointed
  • - The chisel Point is less apt to leave the medullary cavity or penetrate out of the opposite cortex
  • - A pin-chuck is required for its insertion
  • - Due to its round, smooth shape it offers little resistance to rotation
  • - Stability can be improved by use of
  • - Auxillary fixation like K-E app
  • - Or circlage or hemicerclage wires
  • - Or multiple pinning (stack pinning)
  • - Using a large pin that fills the medullary cavity
  • - Placing pin so as it exerts a dynamic force.
  • - Healing is due to the development of periosteal bridging callus
  • - When K-E apparatus is to be used in conjunction with an I.M. pin the size of IN pin should be about 60-70% of the medullary cavity
  • - Should be used in small & medium breeds.
  • - Pins can be used with an open or closed approach
  • - Used mostly in fractures of long bones i.e. femur, Tibia, Humerus, ulna Radius?
  • -- Closed technique used when fracture is of recent origin easily reducible, and stable.

Indications:

  • - Diaphyseal Fr of long bones
  • - Irregular transverse & short oblique Fr.
  • - In spiral & comminuted with auxillary fixation

Advantages:

  • - Simple technique
  • - Inexpensive
  • - Minimal surgical time

Disadvantages:

  • - No rotational stability or resistance to compression or traction forces
  • - Damage to medullary blood vessels & ostrozenic tissue

5.23.13 TECHNIQUE OF I.M. PINNING

5.23.13.1 Retrograde Technique: (Femur) (Humerus)?

  • - Pin is inserted from the fracture site into the medullary canal of the proximal fragment with a steady pressure and back & forth quarter turns of the pin chuck.
  • - The pin is advanced up the medullary cavity till it comes out of the trochanteric fossa and out of skin
  • - The pin chuck is disengaged and applied to the pin protruding from the skin. The pin is withdrawn slowly, again by making 1/4th turn of the wrist, till its end is level with the distal end (Fracture end) of the proximal segment
  • - Fr. is reduced aligned and held in this position
  • - Pin chuck is advanced so as to send the pin down the medullary cavity of the distal segment where it is seated in the metaphyseal epiphyseal area.
  • - The pin setting is evaluated by measuring with an other pin of same length or by radiography
  • - The excess pin, protruding from the skin at trochanteric fossa is cut with pin cutter as close as possible and pushed under the skin
  • - Extreme caution is taken to avoid penetration of distal/adjacent joint surface.

5.23.13.2 Antegrade/Normograde/Direct pinning:

  • - The pin is directed into the medullary canal of one fragment from the appropriate point on the epiphyseal area up to the fracture line
  • - The fracture is reduced and held in reduction while the pin is advance across the fracture line into the other segment and seated securely

Point of Start/Entrance

Bone

Seating of pins

Subtrochanteric fossa

Femur

Distal caudal condyle

Ant crest of greater tubercle

Humerus

Medial condyle

Medial slightly behind straight

Tibia

Medial malleolus

patellar ligament

Ulna

Olecranon process

Distal epiphysis at thecranial border

Radius

(Mostly avoided)

5.23.13.3 Multiple (Stacking) Pins:

  • - Provide more secure fixation
  • - Pins can be inserted retrograde/normograde
  • - Or first pin retrograde & then following reduction additional pins antegrade multiple points or contact provide more rigid fixation against rotational stress
  • - However, tendency of pins to migrate
  • - Caution against splitting the none when cortex is thin or there are fissures present

Cross Pins: Pins are placed across fracture line at converging angles so as to provide 2 or more points of fixation.

  • - Pins should not cross at Fr. line

Indication: Metaphyseal - Epiphyseal Fracture I-IV (Salter)

Advantage: Provides stable reduction without invading joint.

Early joint motion

Disadvantage: - Retard physeal growth?

  • - More difficult than direct pinning
  • - Pin retrieval difficult

5.23.13.4 Rush pinning

  • - Pin when driven into bone provides three point fixation. As in a spring loaded tension
  • - Rush pin has a bevel at one end (to slide off the far cortex and glide along inner cortical surface) and a hook on the other (to engage in the cortex)
  • - Requires a guide hole to insert the pins
  • - Pins come in 3/32", l/8", 3/16" & 1/4" diameter
  • - Steinmann wires/pins can be made into or used as rush pins or in rush pin principle
  • - 2 pins are inserted into metaphyseal region at an acute angle of 30 to the long axis of bone
  • - Pins are hammered alternately slowly in and up the medullary cavity. The beveled end of the pin bounces of the opposite endosteum to engage or rest against the starting cortex. This provides three point fixation.

Indication: Fr. of metaphyseal region Fr. of Proximal humerus & Tibia and distal femur

Disadvantage: Premature closure of physis, if used in young animals?

5.23.14 CLASSIFICATION OF PLATES (per function)

5.2314.1 Compression plate:

used as static Intra frag. compression or dynamic compression (Tension band)

Plate must have sufficient rigidity for the size & weight of the animalShould be applied on the tension side of the bone to achieve dynamic compression

Compression can be achieved by the use of DCP plate, external tension device, semitubular plate

Indications for the use of DCP plate

  • - Transverse fracture
  • - Short oblique Fr,
  • - Osteotomies
  • - Arthrodeses
  • - Non-unions

Principle:

A DCP plate has oval gliding holes. When are drilled by the use of a loading drill guide, it helps make the holes in the bone towards the top of the bone plate glide holes thereby giving compression of about l mm when screws are tightened.

Application of a 3.5mm DCP to a Tr. Fr.- an example

  • - Contour an aluminum template to the surface of the reduced bone
  • - Form the 3.5 mm dynamic compression plate according to the shape of the template
  • - Over bend the plate in its unperforated middle part so that it will be very slightly raised from the bone
  • - Drill the first hole 7-8 mm from the fracture gap, through both cortices using the 2.0/2.5 mm drill bit through a drill guide (load) with eccentrically situated hole and an arrow pointing towards the fracture gap

Measure the depth (length) of the hole (i.e.Width of the bone) by depth gauge

  • - Tap both cortices with 3.5 mm tap through a tap sleeve
  • - While keeping the Fr. under the plate in reduced position tighten the 3.5 screw through the plate to the bone
  • - Similarly tighten the other first screw across the Fr line in the opposite segment
  • - This should reduce the Fr gap by 2mm
  • - Apply the remaining screws in both segments with a green ventral 3.5 mm drill guide

Green/Neutral drill guide provides 0.1 mm compression for each screw tightened

Retighten all the screws Check again if any motion is seen in any screw Screws are tightened by finger pressure only

5.23.14.2 Neutralization plate:

Splinting of a lag scretw fixation as screws alone provide a very questionable stability to fracture area

Plate neutralizes and transmits weight bearing forces that would set in the fracture area

Indications:

Spiral, oblique & comm. diphyseal Fr.

Application:

Same as in compression plating except the holes are drilled using neutral guide Plate is bent precisely to the contour of the bone. A DCP (or round hole plate) can be used.

C. Buttress Plate:

  • - used to shore up a fragment of bone Fr. bridge or splint a Fr. area to maintain leg length

Indications:

  • - comminuted diaphyceal fractures
  • - Proximal tibia1 plateau Fr.
  • - Osteotomies for leg lengthening procedure
  • - Some metaaphyseal Fr.
  • - Where distraction-support is needed

Application:

Plate fixed to main fragments with primarily the objective of holding them the correct distance apart while reconstructing the bone shape. If DCP is used the screws are seated using 2.0 mm drill sleeve inserted through the plate hole towards the fracture. A11 screws are inserted in buttress position

- Application of screws (the steps involved) is same as with other plating techniques

 

5.24 BONE PLATES

5.24 BONE SCREWS

5.24.1 Cortical: -

used primarily in the diaphyseal bone

  • - Screws are fully threaded with more threads per unit length than cancellous screws.
  • - Threads are shallower & more flat pitched
    • Cancellous: used to compress fragments of epiphyseal & metaphyseal area
  • - Maybe partially or completely threaded
  • - few threads per unit length
  • - Threads are deep & relatively high pitched

Indication:

a) Primary fixation alone in epiphyseal or metaphyseal fractures

  • - Cancellous screws; partially threaded are implanted with no threads crossing the fracture line after drilling the appropriate size hole in both segments, tapping it and the seating the appropriate length screw .
  • - Some cancellous screws are self tapping
  • - Tightening the screw, brings compression of the segments as the near fragment slips on the smooth shank

5.24.2 Intrafragmentary Compression with a cortical screw

  • - Insert in a lag fashion or to achieve a lag effect
  • - It requires drilling a larger hole (the diameter equal to the size of the screw) in the near cortex and another hole equal to the size of the core of the screw in opposite or far c o r t e x
  • - The far cortex is tapped with a corresponding tap, diameter of which is equivalent to the size or the screw to be implanted
  • - The tap is always moved 2-3 rotation forward & 1/4 rotation backwards

An appropriate size ( ie . length ) screw is tightened

Steps involved

  • i) drill near cortex with a drill bit the size of the screw
  • ii) drop drill sleeve through it and drill the far cortex with a drill bit corresponding to the screw core size
  • iii) Measure after countersinking
  • iv) Tap with the same size tap as the screw size

    TENSION BAND WIRE:

    Principle:

    • - Active distracting forces are counteracted and converted into compressive forces - It is achieved by inserting 2 K- Wires & a tension band orthopedic wire

    Indication: Avulsion fracture or osteotomies of any apophyseal area - Avulsion Fr. of tibial tubercle, medial malleolus, tuberculosis or acromial process of scapula

    • - Arthrodesis of inter tarsal joints
    • Technique for repairing Fr. of Olecranson Process:
    • - Reduce Fr.
    • - Insert two pins from the top of olecranon process starting from its caudomedial/lateral corners across the fracture line to engage the cranial cortex of the ulna distally. Reep pins parallel to each other
    • - A transverse hole is drilled through the diaphysis below the fracture line & an orthopedic wire is passed through it
    • - The wire is twisted in figure eight fashion and tightened by twisting
    • - Individual K-ires are bent backwards, cut & rotated anteriorly so that ends are burried in soft tissue

5.24.3 Cerclage or Hemicerclage wire:

  • - Circle of wire that completely or partially goes around the circumference of a bone
  • - They are not used as sole method of repair or fixation in any type of fracture

Indications:

  • - Long oblique, spiral & comminuted Fr
  • - For auxillary fixation
  • Technique (General):
  • - Use wires of sufficient strength 10, 20,22 gauge
  • - apply tightly
  • - Used on diaphyseal fractures
  • -- Notching may be needed
  • - Space them about l cm apart & 0.5 cm from Fr. line
  • - Never use less than 2 wires
  • - Length of Fr. line should be double the diameter of bone

5.24.4 Application of Cerclage wire:

  • - reduce fracture segments
  • - pass the wire around both segment with a wire passer or a curved hemostate
  • - Be sure not to include any soft tissue between wire & bone
  • - Twist wire ends uniformly
  • - When it is tight keep tightening & bend it towards bone
  • - Cut excess off

5.24.5 Hemicerclage application:

  • - Drill a hole on either side of Fr. line
  • - Pass a wire through the holes and twist the wire ends together to compress the fracture line
  • - Cut excess wire off
  • - Can be used around an I.M. Pin